The Drying Time Recorder adopts stepmotor, it can provide accurate move control. Six needle clips are installed on the carrier, the needle lowered onto the wet coating moves across the glass strip by the carrier’s shifting. The unit will automatically switch off at the end of test. Raise a pressure plate which is below the back of carrier, reset carrier by hand move. A scale label installed in front of the unit is convenient for the conversion of shift distance to shift time. A total of five different speeds corresponding to 3 ~ 48 hours timer can be chosen.
Main Technical Parameters:
Power: 220V 50Hz
Dimensions of glass panel: 23×340 mm
Five different speeds: 3 / 6 / 12 / 24 / 48 h
Distance of whole journey: 304.8mm (1 foot)
Needles: holding ≤Ф2 dia. with rounded tip
Overall dimensions: 490×220×130 mm(length by width by height)
Weight: 8 kg
Operating Procedure:
1. Preparing work before test:
(1) Make all measurements at constant temperature and humidity.
(2) The instrument should be put on the strong and flat table, far from high magnetic field、 highfield、high electromagnetic pulse.interference
2. Operating Procedure:
(1) Plug in power, switch on.
(2) If the carrier is not in starting position, raise a pressure plate which is below the back of carrier, reset carrier by hand move.
(3) Place needles (with rounded tips) on the needle clip, make needle pole vertical to glass strip.
(4) Loosen holding screw which is on the support frame, take down glass strip. Coat required wet film on the glass strip using the right film applicator
(5) Place the glass strip on the support frame, suitable tighten holding screw which is on one end.
(6) Lower needles gently onto the wet coating.
(7) Select the suitable speed by adjusting the speed switch. Turn the recorder on, needles will be drawn along the glass strip by the carrier. If you need to change choice, push stop button first, then reselect. The unit will automatically switch off at the end of test.
(8) Observe the various stages of film curing and drying, evaluate the results.

Levelling Basic trace Ripped film Surface trace Dry
Explanation: See figure, the film starts to tack-free after doing test 2 hours ( stage 3 starting position );
Not reach dry-through until 5 hours ( stage 5 starting position, mark disappears ).
NOTE:
Guide screw and slide rod should be lubricated from both side of open-end after using a period of time.
Outline drawing:
Film Applicator and holder(order separately):

Ordering Information:
| Type | Name |
| GZY-2 | Line Drying Time Recorder |

Zahn cup อุปกรณ์วัดความหนืด (conforms to ASTM D4212) 


รีวิว
ยังไม่มีบทวิจารณ์